IEEE Transactions on Biomedical Engineering · 1980 · 22 citations · 14 references
MedicineBioengineeringBiomedical EngineeringBiomedical Signal ProcessingUltrasonographyUltrasoundDiagnostic UltrasonographyDoppler UltrasonographyDoppler UltrasoundUltrasonicsUltrasound ImagingUltrasound InstrumentationBlood Flow Doppler AssessmentDoppler ArtifactsDoppler Blood Flow AssessmentFast Fourier TransformReal-time Data AnalysisThree-dimensional Doppler UltrasonographyTime ResolutionVascular Flow Velocity Measurement
Development of a real-time system for three-dimensional display of Doppler shifted frequencies within an ultrasound flowmeter signal is described. The device is based on a fast Fourier transform (FFT) spectrum analyzer which provides both high frequency and time resolution over a variable range of frequencies up to 40 kHz. Signal preprocessing filters unwanted low frequency artifacts due to arterial wail motion and automatically controls input gain to limit peak amplitudes while matched phase shift networks translate forward and reverse components for a directional output. A digital first moment is calculated and the display may be viewed in a continuous or freeze frame mode. Clinical applications underway have shown improved detail and accuracy in the use of ultrasound for diagnosing various vascular abnormalities.
14
Cardiovascular applications of ultrasound
P.N.T. Wells · Ultrasound in Medicine & Biology · 1975 · 309 citations
Medicine, Cardiology, Radiology +12
Detection of carotid occlusive disease by ultrasonic imaging and pulsed Doppler spectrum analysis.
Blackshear Wm, Phillips Dj, Thiele Bl et al. · PubMed · 1979 · 208 citations
Medicine, Cardiology, Radiology +16
Noninvasive ultrasonic carotid angiography: prospective validation by contrast arteriography.