Mean values (n=10 drill sequences) for maximum recorded temperature (Max TºC), change in temperature (ΔTºC) from baseline were as follows. The changes in temperature (ΔTºC) were 1.57ºC and 2.46ºC for the lowest and the highest values, respectively. Drilling at 50 rpm without irrigation did not produce overheating. There was no significant difference in heat production between the 3 implant drill systems (p>0.05). No implant drill system produced heat exceeding 47ºC, which is the critical temperature for bone necrosis during low-speed drilling. Low-speed drilling without irrigation could be used during implant site preparation.