The study introduces radiofrequency pulses that enable spatially periodic excitation, inversion, and refocusing. These pulses are generated by multiplying existing RF pulses with a Dirac comb or via Fourier series expansion, implemented in a spin‑echo sequence, and optimized for fixed pulse duration. The pulses achieve power deposition independent of slice number, reducing per‑slice SAR, while delivering excellent image quality in phantoms and human head, and promise increased scanning efficiency at high field strengths. Published in Magn Reson Med 2011; © 2011 Wiley Periodicals, Inc.

Abstract

Abstract This communication describes radiofrequency pulses capable of performing spatially periodic excitation, inversion, and refocusing. The generation of such pulses either by multiplication of existing radiofrequency pulses by a Dirac comb function or by means of Fourier series expansion is described. Practical schemes for the implementation of such pulses are given, and strategies for optimizing the pulse profile at fixed pulse duration are outlined. The pulses are implemented using a spin‐echo sequence. The power deposition is independent of the number of slices acquired, and hence the power deposition per slice is considerably reduced compared to multislice imaging. Excellent image quality is obtained both in phantoms and in images of the human head. These pulses should find widespread application for multiplexed imaging, in particular at high static magnetic field strengths and for pulse sequences that have a high radiofrequency power deposition and could lead to dramatic increases in scanning efficiency. Magn Reson Med, 2011. © 2011 Wiley Periodicals, Inc.

References

24