Simultaneous assays of glucose, urate, and cholesterol in blood serum by amperometric flow injection analysis

Tao Yao, Masahiro Satomura, Tadaki Nakahara

Electroanalysis · 1995 · 19 citations · 16 references

Concepts

Abstract

Abstract Glucose, uric acid, and cholesterol in serum were assayed simultaneously in a flow injection system by using a 16‐way switching valve and a parallel configuration of three immobilized enzyme reactors. Nafion film‐coated and Nafion/cellulose acetate film‐coated platinum electrodes were used to selectively detect the hydrogen peroxide generated enzymatically in the enzyme reactors, without any interferences from electroactive species such as L ‐ascorbate and proteins present in the serum. The signal current was linearly related to the glucose concentration in the range of 5–700 mg/dL, to the urate concentration in the range of 1–80 mg/dL, and to the cholesterol concentration in the range of 5–100 mg/dL. The precision (coefficient of variation) was better than 2.8% for the simultaneous assay of the three species in human control serum. The assay speed was up to 32 samples per hour.