Measuring lung water: Ex vivo validation of multi‐image gradient echo MRI

Sebastiaan Holverda, Rebecca J. Theilmann, Rui Carlos Sá, Tatsuya Arai, Evan Hall, David J. Dubowitz, G. Kim Prisk, Susan R. Hopkins

Journal of Magnetic Resonance Imaging · 2011 · 40 citations · 13 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: To validate a fast gradient echo sequence for rapid (9 s) quantitative imaging of lung water. MATERIALS AND METHODS: Eleven excised pig lungs were imaged with a fast GRE sequence in triplicate, in the sagittal plane at 2 levels of inflation pressure (5 and 15 cm H(2) O), an intervention that alters T(2) *, but not total lung water. Images were acquired alternating between two closely-spaced echoes and data were fit (voxel-by-voxel) to a single exponential to determine T(2) * and water content, and compared with gravimetric measurements of total water. RESULTS: T(2) * averaged 1.08 ± 0.02 ms at 5 cm H(2) O and 1.02 ± 0.02 ms at 15 cm H(2) O (P < 0.05). The measure was reliable (R(2) = 0.99), with an average mean error of 1.8%. There was a significant linear relationship between the two measures of water content: The regression equations for the relationship were y = 0.92x + 19 (R(2) = 0.94), and y = 1.04x + 4 (R(2) = 0.96), for 5 and 15 cm H(2) O inflation pressure respectively. Y-intercepts were not statistically different from zero (P = 0.86). CONCLUSION: The multi-echo GRE sequence is a reliable and valid technique to assess water content in the lung. This technique enables rapid assessment of lung water, which is advantageous for in vivo studies.