Publication | Open Access
The self-calibration of a retarding-type Mott spin polarimeter with a large collection angle
32
Citations
10
References
2006
Year
EngineeringMeasurementMagnetic ResonanceEducationLarge Collection AngleSecondary ElectronsSpin DynamicSpin PhenomenonCalibrationEffective Sherman FunctionInstrumentationPrecision MeasurementPhysicsSynchrotron RadiationPolarization ImagingSpintronicsApplied PhysicsDynamic Nuclear PolarizationNeutron Scattering
We have developed a compact retarding-type Mott spin polarimeter with a large collection angle at the Hiroshima Synchrotron Radiation Center and obtained a reliable value of the effective Sherman function by a self-calibration method. The spin polarization of secondary electrons was evaluated by extrapolation of the left-right scattering asymmetry at the zero energy-loss limit, which enabled us to obtain the effective Sherman function (Seff) and the figure of merit of our polarimeter at different operating conditions.
| Year | Citations | |
|---|---|---|
Page 1
Page 1