Concepedia

Publication | Open Access

Constrained Cups Appear Incapable of Meeting the Demands of Revision THA

86

Citations

50

References

2011

Year

Abstract

Failure of the locking liner ring and loosening of the acetabular cup are the primary causes of mechanical failure with constrained liners; polyethylene is an inadequate material for restricting motion of the hip to prevent instability. The durability of these devices is unlikely to improve unless the mechanical demands are modified through increased range of motion leading to less frequent rim impingement.