Monitoring the NSL Standard of EMF Using the AC Josephson Effect

I K Harvey, J.C. Macfarlane, R. B. Frenkel

Metrologia · 1972 · 28 citations · 13 references

Concepts

Abstract

The techniques used for the monitoring of the NSL standard of emf using the ac Josephson effect are reported. The standard has been monitored since 29th May 1970 and the reported measurements span a period of 482 days. These measurements have predominantly used niobium point contact Josephson junctions operated at the 1 mV level.

References

13