Review of Scientific Instruments · 2012 · 10 citations · 5 references
MedicineElectrical EngineeringNuclear EngineeringNuclear InstrumentationRadiation DosimetryRadiation PhysicsAccelerator PhysicsElectron Beam PhysicsRadiation TherapyBeam Pulse DurationElectron AcceleratorsElectron Beam AccelerationPulse RadiolysisRise TimeVoltage
A high-voltage pulse-slicer unit with variable pulse duration has been developed and integrated with a 7 MeV linear electron accelerator (LINAC) for pulse radiolysis investigation. The pulse-slicer unit provides switching voltage from 1 kV to 10 kV with rise time better than 5 ns. Two MOSFET based 10 kV switches were configured in differential mode to get variable duration pulses. The high-voltage pulse has been applied to the deflecting plates of the LINAC for slicing of electron beam of 2 μs duration. The duration of the electron beam has been varied from 30 ns to 2 μs with the optimized pulse amplitude of 7 kV to get corresponding radiation doses from 6 Gy to 167 Gy.
5
One-electron reduction of thionine studied by pulse radiolysis
S.N. Guha, P.N. Moorthy, Kamal Kishore et al. · Journal of Chemical Sciences · 1987 · 303 citations
High-speed gated x-ray imaging for ICF target experiments (invited)
D. K. Bradley, P. M. Bell, J. D. Kilkenny et al. · Review of Scientific Instruments · 1992 · 111 citations