Development of high-voltage pulse-slicer unit with variable pulse duration for pulse radiolysis system

J. Upadhyay, Mrinalini Sharma, C. P. Navathe, Madhu Toley, Suhashine Shinde, Shirish A. Nadkarni, S.K. Sarkar

Review of Scientific Instruments · 2012 · 10 citations · 5 references

Concepts

Abstract

A high-voltage pulse-slicer unit with variable pulse duration has been developed and integrated with a 7 MeV linear electron accelerator (LINAC) for pulse radiolysis investigation. The pulse-slicer unit provides switching voltage from 1 kV to 10 kV with rise time better than 5 ns. Two MOSFET based 10 kV switches were configured in differential mode to get variable duration pulses. The high-voltage pulse has been applied to the deflecting plates of the LINAC for slicing of electron beam of 2 μs duration. The duration of the electron beam has been varied from 30 ns to 2 μs with the optimized pulse amplitude of 7 kV to get corresponding radiation doses from 6 Gy to 167 Gy.

References

5