Archives of Ophthalmology · 1989 · 94 citations · 5 references
MedicineOphthalmologySurgeryClinical Vision ScienceOcular AnatomyOphthalmic SurgeryCorneal AnatomyCorneal TopographyAnterior Corneal CurvatureCorneal Curvature MeasurementCorneal Optical ModelingCorneal Topography Assessment
Surgical manipulation of corneal shape requires an accurate and precise method of measuring anterior corneal curvature from apex to limbus. We evaluated the accuracy and precision of the Bausch & Lomb Keratometer, Kera Corneascope, and Computed Anatomy Corneal Modeling System by measuring comparable locations on four calibrated steel test balls. The Keratometer and Corneal Modeling System were more accurate and precise than the Corneascope. No statistically significant difference was found between the Keratometer and Corneal Modeling System with respect to accuracy or precision. In addition, the accuracy and precision of all 31 rings of the Corneal Modeling System were studied. Rings 2 through 26 were read accurately and precisely within +/- 0.25 diopter on three of the four balls. Values on the steepest ball were within +/- 0.37 diopter.
5
J. James Rowsey · Archives of Ophthalmology · 1981 · 130 citations
Medicine, Ophthalmology, Surgery +17
Method for Calculation of Corneal Profile and Power Distribution
J.D. Doss, R. L. Hutson, J. James Rowsey et al. · Archives of Ophthalmology · 1981 · 93 citations
A new photogrammetric method for quantifying corneal topography.
Kenneth L. Cohen, Nancy K. Tripoli, Alvin C. Pellom et al. · PubMed · 1984 · 38 citations
Corneoscopy in Keratorefractive Surgery
J. James Rowsey, Michael S. Isaac · Cornea · 1983 · 28 citations
Medicine, Ophthalmology, Surgery +17