The Journal of the Acoustical Society of America · 1967 · 90 citations · 0 references
EngineeringMechanical EngineeringMetrologyUltrasoundVibrationDisplacement MeasurementInterferometryMode ShapesTransducer TechnologyUltrasonicsVibration MeasurementDisplacementInterferometric Displacement MeasurementInterferometric MeasurementLaser Displacement MeasurementLaser InterferometryMode Shape AnalysisVibration Displacement MeasurementVibration Mode Shapes
Four experimental methods are described in a tutorial way for use with a laser interferometer in measuring vibrational displacements and mode shapes of a transducer face. The relative merits and experimental difficulties of the methods are discussed. Quantitative data to show the type of experimental agreement that one might expect between the various methods and two illustrative mode shape measurements are given. In general, by proper choice of one or more of the methods, displacement measurements of magnitude 0.1–6000 Å at sonic and ultrasonic frequencies can be made at a point 0.1 mm in diameter. Mode shapes are determined simply by moving the laser beam a measured distance between points. Displacement data presented show a random error for all methods of 5% in the range 0.5–80 Å and less than 5% above 80 Å.