Publication | Closed Access
Analysis of retrieved implants: Crystallinity changes in ultrahigh molecular weight polyethylene
57
Citations
10
References
1982
Year
The extent of in vivo induced structural alterations were evaluated in 13 retrieved implants ranging from two weeks' to seven years' implantation. A statistically significant difference in crystallinity was seen between weight-bearing and non-weight-bearing regions of the implants. Further, in the weight-bearing region, crystallinity was higher in components implanted for two or more years than in those implanted for less than one-half year. These results were consistent with our prior in vitro study on the effects of sterilization and aging. Taken together, the data indicate that UHMW polyethylene is not a static material but is continually dynamic changes in vivo.