Annals of Clinical Biochemistry International Journal of Laboratory Medicine · 1991 · 13 citations · 16 references
Life SciencesNephrologyBioavailabilityClinical ImmunologyImmunotechnologyEnzyme-linked Immunosorbent AssayImmunoassay DevelopmentProtein BindingAssay SpecificityDetection Limit AssessmentDetection Limit EstimationImmunoassay MethodologyMethod Detection LimitRetinol-binding Protein AssessmentSerum Retinol Levels
We describe a simple, rapid and sensitive homogeneous immunoassay for urinary retinol-binding protein (RBP) using latex particle-enhanced turbidimetric immunoassay. Rabbit anti-human RBP is covalently coupled to 40 nm latex particles and the assay performed on the IL Monarch 2000 centrifugal analyser, with a 20 microL sample volume and the reaction monitored at 340 nm over an 8 min period. The assay range is 0-6 mg/L with a detection limit of 25 micrograms/L. The within and between assay coefficients of variation are less than 1.5% and less than 2.5%, respectively. Comparison with radioimmunoassay for RBP showed good agreement.
16
Assessment of urinary retinol-binding protein as an index of proximal tubular injury.
A. Bernard, A Vyskocil, P Mahieu et al. · Clinical Chemistry · 1987 · 188 citations · Full text
Pharmacy, Nephrology, Toxicology +17
Current concepts in proteinuria.
K V Waller, Kory M. Ward, John D. Mahan et al. · Clinical Chemistry · 1989 · 142 citations