Review of Scientific Instruments · 2004 · 25 citations · 3 references
EngineeringNuclear EngineeringInstrumentationNuclear InstrumentationPlasma PhysicsDetector TechnologyInterferometryOptical InstrumentationRadiation InstrumentationTemperature MeasurementElectron DensityElectron Density DistributionsLaser InterferometryPolarimeter DesignPolarimetric Instrument DesignPolarimetryPolarimetry InstrumentsTime Resolution
In order to measure the line integrated electron density on the International Thermonuclear Experimental Reactor (ITER), a vibration-compensated five-channel CO2 laser interferometer/polarimeter system is being developed. The design of the interferometer/polarimeter has been improved from the previous design, which used a CO2 laser (wavelength, λ=10.6 μm) and a CO laser (λ=5.3 μm), on the basis of experience gained with a dual CO2 system (λ=10.6 and 9.27 μm) on JT−60U. The JT−60U system gives good reliability for both interferometry and polarimetry operation. The performance of the dual CO2 laser is expected to meet the requirements for measurements on ITER (accuracy; 1%, time resolution; 1 ms).