Publication | Closed Access
Implantation of a custom intraocular lens with a 30-diopter torus for the correction of high astigmatism after penetrating keratoplasty
36
Citations
9
References
2003
Year
We present a 62-year-old woman who had implantation of an intraocular lens (IOL) with a torus of 30.0 diopters (D) in an eye after penetrating keratoplasty. The patient presented with cataract and a best corrected visual acuity of 20/400. Penetrating keratoplasty was performed in 1975 because of decompensated keratoconus. With a clear but inferiorly decentered corneal graft, visual acuity was limited because of keratoplasty-related high corneal astigmatism of 22.5@156 that could not be successfully treated with spectacles or contact lenses. A custom toric IOL with -30.0 D of cylindrical power was implanted in the capsular bag via a sclerocorneal tunnel incision. Additionally, a spherical IOL of -15.0 D sphere was implanted in the ciliary sulcus. Six months after implantation, the IOLs were still well centered and the uncorrected visual acuity was 20/60.