Publication | Closed Access
Electrically Evoked Compound Action Potential Amplitude Growth Functions and HiResolution Programming Levels in Pediatric CII Implant Subjects
62
Citations
21
References
2004
Year
In pediatric subjects with the Clarion HiFocus electrode, ECAP growth function thresholds appear to decrease with the presence of the Silastic positioner but are unaffected by electrode position along the array. Growth function slope, however, depends on electrode position along the array but not on the presence of the positioner. ECAP programming levels can reliably predict stimulus intensities within the patients' dynamic ranges, but considerable variability is seen between ECAP thresholds and HiRes programming levels.