Transplant Infectious Disease · 2006 · 30 citations · 6 references
Life SciencesImmunologyInfectious DiseasesClinical ImmunologyImmunology Of Infectious DiseasesTransplant ImmunologyAntigen Detection AssaysAutoantigensAutoimmune DiseasesDiagnostic ImmunologyImmunosuppressionAcquired Immunodeficiency SyndromeAntibodiesAssayDiagnostic AntigensImmunoassaysImmunodiagnostic AssaysRabbit Antithymocyte Globulin
False-positive Histoplasma antigenemia was reported in solid organ allograft recipients who had received rabbit anti-thymocyte globulin (RATG, RATG) caused by human anti-rabbit antibodies (HARA). A second-generation Histoplasma antigen detection assay was developed to overcome false positivity caused by HARA. With the second-generation assay, false-positive results were eliminated in 18 of 19 cases without reduction in the sensitivity in patients with histoplasmosis. In fact, sensitivity for detection of antigenuria in patients with acquired immunodeficiency syndrome and disseminated histoplasmosis was higher in the second-generation assay. Physicians should be aware of the potential for false-positive results in sandwich immunoassays in specimens from patients who have received RATG.
6
Human Anti-Animal Antibody Interferences in Immunological Assays
Larry J. Kricka · Clinical Chemistry · 1999 · 588 citations · Full text
M M Durkin, Patricia Connolly, L. Joseph Wheat · Journal of Clinical Microbiology · 1997 · 106 citations · Full text
Interference by human anti-mouse antibody in two-site immunoassays
Larry J. Kricka, Deborah Schmerfeld‐Pruss, Marilyn B. Senior et al. · Clinical Chemistry · 1990 · 102 citations