Journal of High Resolution Chromatography · 1984 · 19 citations · 8 references
Natural SciencesChemistryAir PollutionAnalytical ChemistryChromatographic AnalysisChromatographyExtraction TechniquesGas ChromatographyInstrumental AnalysisSolvent ExtractionAnalytical Detection Of Organic PollutantsAromatic HydrocarbonsCapillary Gas ChromatographyExtraction Of Organic CompoundsPolycyclic Aromatic Hydrocarbon AnalysisPolycyclic Aromatic Hydrocarbon RemovalPolycyclic Aromatic HydrocarbonsSample Injection
Abstract A commercial thermodesorption and cold trap unit can be used as a large sample volume injection device in high resolution gas chromatography. Volumes of more than 100 μl can be injected. More than 97% of solvents, such as hexane, can be removed by a Tenax TA precolumn without any loss of compounds with volatilities comparable to naphthalene. Experimental conditions, recovery rates of polycyclic aromatic hydrocarbons, and the reproducibility of the method are described in detail. The method was successfully applied for the determination of PAH in air particulate matter extracts.
8
Michael Oehme, Stein Manø, H. Stray · Journal of High Resolution Chromatography · 1982 · 60 citations
Michael Oehme, H. Stray · Fresenius Zeitschrift für Analytische Chemie · 1982 · 58 citations
G. Grimmer, K.‐W. Naujack, Dietmar Schneider · Fresenius Zeitschrift für Analytische Chemie · 1982 · 53 citations