Medical Engineering & Physics · 2002 · 38 citations · 27 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBiomechanicsOrthopedic BiomaterialsOrthopedic BiomechanicsArthroplasty BiomechanicsBone BiomechanicsBone MechanicsHip BiomechanicsImplant BiomechanicsJoint MechanicsCementless Femoral FixationCementless Stem DesignCortical Bone MechanicsFemoral Component DesignFemoral Stem DesignImplant Material MechanicsTitanium Materials
The strain distribution imposed on a femur following a total joint replacement is an important factor, in proximal bone loss due to stress shielding, and long term clinical success. This study investigated how five different design parameters of a cementless titanium femoral prosthesis influenced cortical strains. Test loads were applied and strains were measured with and without an abductor force simulation, using six human cadaveric femora. The cementless design used demonstrated significant calcar loading proximally and a similar strain distribution to the intact femur distally. Implant gross geometry was the major factor in determining the cortical strain distributions under abductor simulation in both axial and torsional loading.
27
Adaptive bone-remodeling theory applied to prosthetic-design analysis
R. Huiskes, Harrie Weinans, H.J. Grootenboer et al. · Journal of Biomechanics · 1987 · 1.2K citations · Full text
Iekuni Oh, William H. Harris · Journal of Bone and Joint Surgery · 1978 · 294 citations