Concepedia

Publication | Closed Access

Pt resistor thermometry and pressure calibration in a clamped pressure cell with the medium, Daphne 7373

224

Citations

7

References

1997

Year

Abstract

Calibration of a Pt resistance thermometer (Netsushin) in magnetic fields and under pressure in the range of 1.5–300 K and below 1.5 GPa is presented. With the pressure medium, olefin olygomers, Daphne 7373, the pressure is continuous at its solidification and the pressure drop from 300 to 4.2 K by 0.15–0.17 GPa is constant, irrespective of the initial clamped pressure at 300 K. The applicability of the thermometer and the medium for precise study in field and pressure at low temperature is discussed.

References

YearCitations

Page 1