The Mechanical Performance of Cervical Total Disc Replacements In Vivo
24
Citations
26
References
2012
Year
Early clinical failures of Prodisc-C CTDRs display surface damage evidence of metal endplate-endplate impingement, most commonly posteriorly. Backside wear was not evident; however, third-body wear was found. Future studies will determine the clinical impact of these predominant modes of wear on long-term metal-on-polyethylene semiconstrained CTDR performance.