Extreme ultraviolet (EUV) interferometry has been successfully performed for the first time at 0.3 numerical aperture (NA). Extensive EUV “at-wavelength” testing including alignment, was performed on a newly created Micro Exposure Tool (MET) optic designed for sub-50-nm EUV lithographic imaging experiments. The two-mirror, 0.3 NA MET is among the highest resolution light-projection lithography tools ever made. Using both lateral shearing and phase-shifting point-diffraction interferometry, the wavefront was measured across the field of view, and the alignment was optimized in preparation for imaging. The wavefront quality reached 0.55nm RMS (λEUV∕24.5) in a 37-term annular Zernike polynomial series, dominated by higher-order spherical aberration. Measurements included calibrations of the interferometer accuracy, assessment of repeatability, and cross-comparisons of visible and EUV interferometric measurements.