Review of Scientific Instruments · 1992 · 13 citations · 1 references
Natural SciencesMaterials ScienceCrystal StructureRadiation MeasurementRadiation PhysicsBeam AnalysisX-ray OpticsBeam InstabilitiesBeam Position MonitoringHeat Load AssessmentHeat Load DistributionHigh-energy Particle RadiationLoad EffectsMisalignment ToleranceX-ray Monochromators
The double-crystal monochromator on beam line 3.4 at the Daresbury synchrotron radiation source covers the energy region 0.8–4 keV, routinely using crystal pairs of beryl, quartz, InSb, Ge, and Si. The beam line has a unique system to cut off high-energy radiation. The whole beam line has been built on a swinging arm, which allows the glancing angle on a chromium coated premirror to be varied between 0.5° and 3°. This not only reduces the heat load on fragile crystals, but also reduces the higher order contributions. We have made a thorough study of the alignment problems arising in the double-crystal geometry. Crystal misalignment, Bragg angular offset, as well as heat-load effects, can cause severe beam movements. A beam-position monitor, which has recently been commissioned, is an invaluable diagnostic tool for these beam movements.
1