Retinal Light Exposure From Operation Microscopes

Joseph L. Calkins, B. F. Hochheimer

Archives of Ophthalmology · 1979 · 75 citations · 13 references

Concepts

Abstract

Several popular surgical microscopes were measured for source radiance. A method has been devised for calculating patient retinal exposure from these instruments. Retinal irradiance is seen to be surprisingly high, assuming dilated pupils and clear media. It ranges from 0.10 to 0.97 W/sq cm. This is one to ten times higher than that produced by an indirect ophthalmoscope. There is evidence linking light exposure during surgery to cystoid maculopathy (Irvine-Gass syndrome).

References

13