Another Complication of Armored Endotracheal Tubes

Kie-Chul Ohn, Wen-hsien Wu

Anesthesia & Analgesia · 1980 · 13 citations · 0 references

Concepts

Abstract

Armored endotracheal tubes are frequently used during anesthesia when bending or compression of the tube is likely to occur as in neurosurgery or head and neck and oral surgery. However, complications may be associated with use of these tubes. They include collapse of the tube inside the inflated cuff, obstruction caused by folding of the inner wall around the connector1,2 or double layering of the cuff preventing deflation.3 The layering phenomenon is aggravated by repeated autoclaving.1 Airway obstruction caused by nitrous oxide infiltration of the tube wall during anesthesia has not been described. The present report depicts this type of complication.