Radiation Protection Dosimetry · 1998 · 47 citations · 0 references
MedicineRadiologyDiagnostic RadiologyRadiation DosimetryRadiation ProtectionBreast AssessmentBreast ExaminationBreast ImagingBreast Radiation CareMammography PhysicsPatient-specific DosimetryBreast Dose CalculationBreast DosimetryDose Management In MammographyMammographyMammography ImagingRadiation Risk In Breast ImagingRadiation Safety In MammographyScreening Mammography
The two methods used to monitor dose in mammography in the UK Breast Screening Programme are reviewed. The first is the measurement of mean glandular dose to a standard breast. The second is the estimation of mean glandular dose for individual women using the X ray exposure factors. Neither method takes account of variations in breast composition and the effect of this on estimates of glandular dose are reported. Applying the standard breast model using PMMA resulted in substantial overestimates of breast doses for women with large compressed breast thicknesses because they tend to have low glandular content. Conversely the use of post-exposure data to estimate the MGDs in women attending screening with large breasts resulted in underestimates of their doses unless the relatively low levels of glandularity were taken into account. When estimating the MGDs for breasts of different sizes, it is important that the models and methods used take account of the strong variation in breast composition with compressed thickness.