Determination of Process-Related Impurities and Degradation Products in Cefaclor by High-Performance Liquid Chromatography

L. J. Lorenz, Fadia N. Bashore, Bernard A. Olsen

Journal of Chromatographic Science · 1992 · 17 citations · 0 references

Concepts

Abstract

A method employing gradient elution high-performance liquid chromatography (HPLC) is described for the determination of process-related impurities and degradation products in the cephalosporin antibiotic, cefaclor. The separation is performed on an ODS column (4.6 mm × 250 mm, 5-μm particles) using 50 mM phosphate buffer, pH 4.0, with a two-stage acetonitrile gradient from 2.25 to 45%. Ultraviolet detection at 220 nm is used. The mobile phase pH is an important factor in the separation of amphoteric compounds present in the sample, with pH 4.0 providing good separation and little fluctuation of retention with changes in pH. Applicability of the method for monitoring cefaclor bulk stability and determining the effect of manufacturing or sourcing changes on quality is demonstrated. The potential for extending the method to formulations is also described.