JAMA · 1970 · 33 citations · 3 references
Health ProfessionsRespiratory TherapyPulmonary Function TestingResuscitationVentilator ManagementCardiopulmonary ResuscitationMechanical VentilationPositive End-expiratory PressureInspiratory Mechanics
Nine commonly used resuscitator bags were connected to an oxygen source, and tested by manually squeezing an 800-cc tidal volume ten times per minute into a test apparatus which measured the oxygen concentration delivered (up to 15 liters of oxygen per minute flowing). They were then tested on conscious volunteers. The Ambu gave the highest oxygen concentration (87% maximum) and the Pulmonator, the lowest (37% maximum). The Ambu with the new E-2 valve was found to be superior in function to the Ambu (with the Ruben valve) and the Ambu (Air Shields). The most common malfunction was sticking of the inflating valve in the inspiratory position. This was found with the Laerdal, Pulmonator, and Ambu (Ruben valve). The Ambu (with E-2 valve) was the easiest to use; the Pulmonator, Aga, and Laerdal (with reservoir), the most clumsy.
3
Automatic Ventilation of the Lungs.
Annals of Internal Medicine · 1970 · 152 citations · Full text
H Ruben · Anesthesiology · 1955 · 43 citations · Full text
APPARATUS FOR RESUSCITATION AND SUCTION
H Ruben, Arne Ruben · The Lancet · 1957 · 19 citations
Health Professions, Emergency Medical Services, Resuscitation +9