A histomorphometric evaluation of screw‐shaped implants each prepared with two surface roughnesses

Ann Wennerberg, Carin Hallgren, Carina Johansson, Stefano Danelli

Clinical Oral Implants Research · 1998 · 341 citations · 0 references

Concepts

TL;DR

The study fabricated 40 screw‑shaped implants with two distinct surface topographies per screw, characterized by confocal laser scanning profilometry (height, spatial, hybrid parameters), and evaluated their histomorphometric integration after 12 weeks in rabbit bone. Blasted surfaces, especially those treated with 75 pm particles (S = 1.4 pm, Sdr = 11.6 pm, Sdr ratio = 1.5), exhibited greater bone contact to the implant surface than turned surfaces.

Abstract

Four different surface modifications were designed. Forty screw‐shaped implants were divided into 4 groups, 10 screws in each. Every screw was prepared with 2 different surface topographies. The surface topography was measured with a confocal laser scanning profilometer and the surface roughness was characterized using 1 height, 1 spatial and 1 hybrid descriptive parameter. After 12 weeks in rabbit bone all screws were histomorphometrically evaluated. Blasted surfaces demonstrated more bone in contact to implant surface compared with turned surfaces. Most bone in close contact to implant surface was found for a surface blasted with 75 pm sized particles, numerically characterized with an average height deviation (S,) of 1.4 pm, an average wavelength (S dr ) of 11.6 pm and a developed surface area 1 ratio (S dr ) of 1.5.