Natural SciencesChemistryToxicologyAnalytical ChemistryEnvironmental ChemistryEnvironmental MonitoringChemical SensorsEnvironmental AnalysisEnvironmental Analytical ChemistryExposure AssessmentExposure MonitoringSulfur ChemistryDetection Limit AssessmentDetection Limit EstimationHydrogen Sulfide ChemistryInstrument Detection LimitInterferenceMethod Detection LimitPersonal Exposure Monitoring
A method for the sampling and subsequent determination of personal exposure to hydrogen sulfide is described. Samples are collected by permeation through a dimethyl silicone membrane into 10 mL of 0.2N NaOH containing EDTA. The collected H2S is converted to a colored product (methylene blue) which is measured spectrophotometrically and the H2S exposure (time-weighted-average) then calculated. The detection limit of the method is 0.01 ppm (0.014 mg/m3) for an 8 h exposure with a working range of 0.1–20 ppm and linear response up to 200 ppm H2S. The device responds in less than 30 sec and is unaffected by variations in humidity and temperature. The only significant interference observed is from chlorine (−6.6%). The use of EDTA increases S= stability as to allow for sampling of over 1 week. A device can be constructed that is small in weight and size, serving as either a personal or area monitor.