Publication | Open Access
Clotting Problems with the Teflon-Silastic Arteriovenous Shunt in Patients on Regular Haemodialysis
11
Citations
6
References
1969
Year
ThrombosisHemodialysisVenous DiseaseVenous PressureMedicineKidney FailureTeflon-silastic Arteriovenous ShuntVascular SurgeryRegular HaemodialysisDialysis TherapySurgeryVascular AccessCoagulopathyChronic Kidney DiseaseNephrologyEmergency MedicineEarly Stage
Episodes of clotting that occurred in 22 patients on regular haemodialysis were studied over a six-month period. The venous pressure during dialysis and the radiology of the Teflon-Silastic arteriovenous shunt were found to be satisfactory guides for the management of the shunt. The failure of the shunt during the early stage was mainly due to technical reasons. Histological study of the excised vessels in removed long-term shunts showed that these had failed because of rigidity and thickening of the vessel wall due to calcium and iron deposits or chronic inflammation, or both.