Development of an amperometric biosensor for phenol detection

L.M.C. Silva, Andréa Medeiros Salgado, Maria Alice Zarur Coelho

Environmental Technology · 2011 · 15 citations · 17 references

Concepts

Abstract

With the aim of searching for an in situ method for monitoring phenol, Agaricus bisporus tissue with tyrosine activity was used as a biocomponent and an oxygen electrode used as a transducer to develop a biosensor. The experimental methodology investigated the relation between dissolved oxygen and phenol concentration using a standard solution. Biosensor calibration was evaluated by studying reaction time and tissue amount necessary to promote a reliable response and to minimize errors. The influence of air saturation of the sample and washing of the electrode was also investigated. Results showed that 5 g of mushroom tissue with a 1 min reaction time promoted the best biosensor response within a phenol concentration range of 5-10 ppm. Washing of the electrode did not change the performance of the analysis; however, initial air saturation caused less variation amongst the samples.

References

17