Performance of Harshaw TLD-100H two-element Dosemeter

11

Citations

3

References

2006

Year

Abstract

One of the advantages of LiF based thermoluminescent (TL) materials is its tissue-equivalent property. The Harshaw TLD-100H (LiF:Mg,Cu,P) material has demonstrated that it has a near-flat photon energy response and high sensitivity. With the optimized dosemeter filters built into the holder, the Harshaw TLD-100H two-element dosemeter can be used as a whole body personnel dosemeter for gamma, X ray and beta monitoring without the use of an algorithm or correction factor. This paper presents the dose performance of the Harshaw TLD-100H two-element dosemeter against the ANSI N13.11-2001 standard and the results of tests that are required in IEC 1066 International Standard.