Clinical Chemistry and Laboratory Medicine (CCLM) · 1988 · 100 citations · 2 references
Life SciencesBiochemistryPhysiologyAnalytical ChemistryCarbohydrate MetabolismContinuous Glucose MonitoringGlycemic MonitoringAcetoacetate MetabolismBlood Glucose RegulationBlood Lactate AssessmentBlood Lactate MeasurementClinical Chemistry AssaysGlucose Analysis In Body FluidsLaboratory Plasma Glucose Testing
Methods are described for the analysis of glucose, lactate, pyruvate, alanine, glycerol, 3-hydroxybutyrate and acetoacetate in perchloric acid extracts of human blood, using the Cobas Bio centrifugal analyser fitted with a fluorimetric attachment. Intra-assay and inter-assay coefficients of variation ranged from 1.9 to 7.9% and from 1.0 to 7.2% respectively. Correlation coefficients ranged from 0.96 to 0.99 against established continuous-flow and manual spectrophotometric methods. All seven metabolites can be measured using a single perchloric acid extract of 20 microliter of blood. The versatility of the assays is such that as little as 100 pmol pyruvate, 3-hydroxybutyrate or as much as 15 nmol glucose can be measured in the same 20 microliter extract.