Blood Glucose, Lactate, Pyruvate, Glycerol, 3-Hydroxybutyrate and Acetoacetate Measurements in Man Using a Centrifugal Analyser with a Fluorimetric Attachment

100

Citations

2

References

1988

Year

Abstract

Methods are described for the analysis of glucose, lactate, pyruvate, alanine, glycerol, 3-hydroxybutyrate and acetoacetate in perchloric acid extracts of human blood, using the Cobas Bio centrifugal analyser fitted with a fluorimetric attachment. Intra-assay and inter-assay coefficients of variation ranged from 1.9 to 7.9% and from 1.0 to 7.2% respectively. Correlation coefficients ranged from 0.96 to 0.99 against established continuous-flow and manual spectrophotometric methods. All seven metabolites can be measured using a single perchloric acid extract of 20 microliter of blood. The versatility of the assays is such that as little as 100 pmol pyruvate, 3-hydroxybutyrate or as much as 15 nmol glucose can be measured in the same 20 microliter extract.