Annals of Plastic Surgery · 2005 · 261 citations · 13 references
MedicineSurgeryAnesthesiaPhysical TherapyVascular SurgeryHemodynamicsNegative Pressure Wound TherapyTissue Perfusion AssessmentBlood FlowBlood Flow DynamicsContinuous Negative Pressure TherapyCutaneous Blood VesselsLaser Doppler ImagingLaser Doppler VelocimetryPolyurethane MaterialsSkin Perfusion Pressure
Vacuum‑assisted closure (V.A.C. Therapy) employs two foams—black polyurethane (PU) and white polyvinyl alcohol (PVA)—with distinct physical properties. The study prospectively evaluates how varying negative pressures and foam types affect cutaneous blood flow in healthy forearm skin. Continuous negative pressure ranging from 25 to 500 mm Hg was applied, and skin blood flow was measured with laser Doppler probes embedded in the foam.
In Brief Vacuum-assisted closure (V.A.C. Therapy) uses 2 distinct types of foams, with different physical characteristics: the black polyurethane (PU) foam and the white polyvinyl alcohol (PVA) foam. This prospective, randomized study evaluates the response of cutaneous blood flow (CBF) in healthy intact forearm skin to varying V.A.C. Therapy negative pressures and both foam types. Continuous negative pressure was used in the range of 25–500 mm Hg. Skin blood flow was measured with noninvasive laser Doppler probes incorporated into the foam. Significant increase in CBF was found with both foams up to negative pressure of 300 mm Hg, with over 5-fold increase (mean: 5.57; SD: 3.32) with the PU foam and nearly 3-fold increase (mean: 2.87; SD: 1.29) with the PVA foam. Comparison of blood flow at baseline and at a negative pressure of 300 mm Hg showed a statistically significant difference (P < 0.001). No decrease in blood flow below baseline was observed during the experiments. A prospective randomized study comparing the 2 VAC foam types demonstrated that the increase in skin blood flow beneath the black polyurethane foam was 67% greater than that beneath the white polyvinyl alcohol foam at b300 mm Hg.
13
Guidelines for measurement of cutaneous blood flow by laser Doppler flowmetry
Andreas Bircher, E. M. de Boer, Tove Agner et al. · Contact Dermatitis · 1994 · 255 citations