Medical Engineering & Physics · 2013 · 21 citations · 41 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBone GraftingOrthopedicsBone Defect ReconstructionBone MechanicsBone ReconstructionBone Growth PlatesFinite Element Analysis Of BoneFinite Element Modeling Of BonePlate Fixation Mechanics
INTRODUCTION: The strengthening effect of prophylactic internal fixation (PIF) with a bone plate at the radial osteocutaneous flap donor site has previously been demonstrated using the sheep tibia model of the human radius. This study investigated whether a finite element (FE) model could accurately represent this biomechanical model and whether stress or strain based failure criteria are most appropriate. METHODS: An FE model of an osteotomised sheep tibia bone was strengthened using 4 types of plates with unilocking or bicortical screw fixation. Torsion and 4-point bending simulations were performed. The maximum von Mises stresses and strain failure criteria were studied. RESULTS: The strengthening effects when applying stress failure criteria [factor 1.76-4.57 bending and 1.33-1.80 torsion] were comparable to the sheep biomechanical model [factor 1.73-2.43 bending and 1.54-2.63 torsion]. The strongest construct was the straight 3.5mm stainless steel unilocking plate. Applying strain criteria the strongest construct was the straight 3.5mm stainless DCP plate with bicortical screw fixation. CONCLUSIONS: The FE model was validated by comparison with the sheep tibia model. The complex biomechanics at the bone-screw interface require further investigation. This FE modelling technique may be applied to a model of the human radius and other sites.
41