Validation of a portable nitric oxide analyzer for screening in primary ciliary dyskinesias

Amanda Harris, Esther Bhullar, Kerry Gove, Rhiannon Joslin, Jennifer Pelling, Hazel Evans, Woolf T. Walker, Jane S. Lucas

BMC Pulmonary Medicine · 2014 · 39 citations · 18 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: Nasal nitric oxide (nNO) levels are very low in primary ciliary dyskinesia (PCD) and it is used as a screening test. METHODS: We assessed the reliability and usability of a hand-held analyser in comparison to a stationary nitric oxide (NO) analyser in 50 participants (15 healthy, 13 PCD, 22 other respiratory diseases; age 6-79 years). Nasal NO was measured using a stationary NO analyser during a breath-holding maneuver, and using a hand-held analyser during tidal breathing, sampling at 2 ml/sec or 5 ml/sec. The three methods were compared for their specificity and sensitivity as a screen for PCD, their success rate in different age groups, within subject repeatability and acceptability. Correlation between methods was assessed. RESULTS: Valid nNO measurements were obtained in 94% of participants using the stationary analyser, 96% using the hand-held analyser at 5 ml/sec and 76% at 2 ml/sec. The hand-held device at 5 ml/sec had excellent sensitivity and specificity as a screening test for PCD during tidal breathing (cut-off of 30 nL/min,100% sensitivity, >95% specificity). The cut-off using the stationary analyser during breath-hold was 38 nL/min (100% sensitivity, 95% specificity). The stationary and hand-held analyser (5 ml/sec) showed reasonable within-subject repeatability(% coefficient of variation = 15). CONCLUSION: The hand-held NO analyser provides a promising screening tool for PCD.