Analysis of partial volume effects on arterial input functions using gradient echo: A simulation study
52
Citations
34
References
2009
Year
Medical UltrasoundEngineeringBiomedical EngineeringMagnetic Resonance ImagingBlood FlowArtery OrientationsGradient EchoAbsolute Blood FlowNeurologyCardiologyBlood Flow MeasurementRadiologyCardiovascular ImagingVascular ImageMedical ImagingMedicineQuantitative Perfusion MetricsNeuroimagingUltrasoundMedical Image ComputingCerebral Blood FlowBiomedical FlowPartial Volume EffectsPhysiologyBiomedical ImagingArterial Input FunctionsStroke
Absolute blood flow and blood volume measurements using perfusion weighted MRI require an accurately measured arterial input function (AIF). Because of limited spatial resolution of MR images, AIF voxels cannot be placed completely within a feeding artery. We present a two-compartment model of an AIF voxel including the relaxation properties of blood and tissue. Artery orientations parallel and perpendicular to the main magnetic field were investigated and AIF voxels were modeled to either include or be situated close to a large artery. The impact of partial volume effects on quantitative perfusion metrics was investigated for the gradient echo pulse sequence at 1.5 T and 3.0 T. It is shown that the tissue contribution broadens and introduces fluctuations in the AIF. Furthermore, partial volume effects bias perfusion metrics in a nonlinear fashion, compromising quantitative perfusion estimates and profoundly effecting local AIF selection.