Medical Physics · 1998 · 29 citations · 7 references
Computed TomographyEngineeringMeasurementPhilips Ac3Image SensorX-ray ImagingRadiographyCalibrationCt ScanPhoton-counting Computed TomographyInstrumentationRadiation ImagingNuclear MedicineRadiologyHealth SciencesLuminescence Noise PowerMedical ImagingDigital ImagingRadiographic ImagingSignal ProcessingElectronic ImagingLuminescence Noise
A comprehensive investigation of the physical imaging performance of a Philips AC3 computed radiography system using fifth-generation image plate technology has been undertaken. Measurements include characteristic response, modulation transfer function (MTF) and Wiener spectra (WS) for standard and high resolution image plates sampled at 10 pixels/mm. These results were used to calculate noise equivalent quanta (NEQ) and detective quantum efficiency (DQE) spectral descriptions of system performance. Luminescence noise and x-ray quantum noise components were separated. From an estimate of the luminescence noise power, the average system gain was calculated and results show a substantial improvement over earlier generations of computed radiography systems for standard image plates.
7
Unified SNR analysis of medical imaging systems
Robert F. Wagner, David G. Brown · Physics in Medicine and Biology · 1985 · 362 citations · Full text
DQE(<i>f</i>) of four generations of computed radiography acquisition devices
James T. Dobbins, David L. Ergun, Lois Rutz et al. · Medical Physics · 1995 · 306 citations
Imaging performance of a digital storage phosphor system
Walter Hillen, U. Schiebel, T. Zaengel · Medical Physics · 1987 · 151 citations