Radiology · 1998 · 34 citations · 0 references
MedicineRadiologyDiagnostic ImagingDiagnostic RadiologySurgical InstrumentationBiopsy TechniquesFluoroscopy-guided ProceduresImage-guided BiopsyNeedle HoldersBiopsy Needle PlacementBiopsy NeedlesCoaxial Biopsy TechniqueFluoroscopic GuidanceFluoroscopy-guided BiopsyGuided Core Needle BiopsyNeedle Aspiration BiopsyNeedle Placement Accuracy
The authors evaluated accuracy and success rates for placement of fine-aspiration, core, and coaxial biopsy needles in pork meat with three needle guide devices and computed tomographic fluoroscopic guidance. Accuracy and reliability with a metallic sponge-forceps needle holder was equal to or greater than those with other devices, and it was preferred by operators due to its lighter weight and single-handed manipulation. All needle holders functioned poorly with thin-walled needles.