Scintillating-fiber 14 MeV neutron detector on TFTR during DT operation

G. A. Wurden, R.E. Chrien, Cris W. Barnes, W.C. Sailor, A. L. Roquemore, Martin Lavelle, P. M. O’Gara, R. J. Jordan

Review of Scientific Instruments · 1995 · 50 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

A compact 14 MeV neutron detector using an array of scintillating fibers has been tested on the TFTR tokamak under conditions of a high gamma background. This detector uses a fiber-matrix geometry, a magnetic field-insensitive phototube with an active HV base and pulse-height discrimination to reject low-level pulses from 2.5 MeV neutrons and intense gammas. Laboratory calibrations have been performed at EG&G Las Vegas using a pulsed DT neutron generator and a 30 kCi 60Co source as background, at PPPL using DT neutron sources, and at LANL using an energetic deuterium beam and target at a tandem Van de Graaff accelerator. During the first high-power DT shots on TFTR in December 1993, the detector was 15.5 m from the torus in a large collimator. For a rate of 1×1018 n/s from the tokamak, it operated in an equivalent background of 1×1010 gammas/cm2/s (∼4 mA current drain) at a DT count rate of 200 kHz.