Publication | Closed Access
Accuracy and Reliability of the Keeler Pulsair EasyEye Non-Contact Tonometer
46
Citations
27
References
2008
Year
The Pulsair EasyEye noncontact tonometer is considered an accurate reliable method in the normotensive population studied but because of a systematic bias in a small number of subjects, it cannot be used interchangeably with the Goldmann tonometer.