Optics Express · 2013 · 57 citations · 18 references
EngineeringBiomedical EngineeringChemistryBiomedical ImagingDiagnostic ImagingMolecular ImagingImagingMedical ImagingMedical Imaging InstrumentationMedical Imaging PhysicsRadiation Imaging DetectorsRadiographic ImagingCoded Aperture ImagingDetector Energy ResolutionEnergy-resolving Detector ArraysMomentum TransferNon-destructive ExaminationNon-destructive TestingQuality InspectionX-ray Imaging Physics
We demonstrate a technique for measuring the range-resolved coherent scatter form factors of different objects from a single snapshot. By illuminating the object with an x-ray pencil beam and placing a coded aperture in front of a linear array of energy-sensitive detector elements, we record the coherently scattered x-rays. This approach yields lateral, range, and momentum transfer resolutions of 1 mm, 5 mm, and 0.2 nm⁻¹, respectively, which is sufficient for the distinguishing a variety of solids and liquids. These results indicate a path toward real-time volumetric molecular imaging for non-destructive examination in a variety of applications, including medical diagnostics, quality inspection, and security detection.
18
Pixellated Cd(Zn)Te high-energy X-ray instrument
P. Seller, S. Bell, Robert J. Cernik et al. · Journal of Instrumentation · 2011 · 112 citations