The Journal of the Acoustical Society of America · 1948 · 24 citations · 0 references
Natural SciencesPhysicsAcoustic DesignAcousticsBuilding PhysicsAcoustic MeasurementRoom AcousticsAcoustic Source CalibrationElectroacoustic MeasurementElectroacoustic SystemsField CalibrationReciprocity CalibrationRoom Acoustics MeasurementTransducer Characterization
A procedure for obtaining a free field reciprocity calibration of a W. E. 640AA transducer in the frequency range 1–100 kc is described. The basis for the calibration is investigated in some detail, particularly with regard to (a) the validity of electroacoustic reciprocity relations applying to the transducer and (b) the nature of the acoustic calibrating field. The transducers were found to be reciprocal. Factors affecting the acoustic field are the degree of approximation of the source to a point transducer, the effect of walls of the anechoic room, and the attenuation in the air. It is shown that when the separation of the transducers used in these measurements is 30 cm, the transducers are essentially points, the room produces very little effect, but the attenuation in the air can produce a significant error at high frequencies unless correction is made for it. The results of the calibration of two W. E. 640AA transducers are presented, giving their responses as speakers and microphones, the latter with and without a protective grid.