The Ternary Coppersilver Selenide—A New Homogeneous Solid State Ion-Selective Electrode for Copper (II)

M. Neshkova, J. Havas

Analytical Letters · 1983 · 16 citations · 17 references

Concepts

Abstract

Abstract The performance characteristics of a copper (II) ion-selective electrode, based on pressed-pellet membrane of ternary CuAgSe, are reported. The sensing material is isostructural with the natural mineral β -eucairite and shows high corrosion resistivity which results in stable and reproducible electrode performance upon ageing. The electrodes exhibit a linear Nernstian response dawn to 5.10−8 M in non buffered medium and down to 10−13M in copper-glycine ion buffer. Data on the electrode stability for a period of two and a half years, on the electrode response time, pH-dependence, selectivity etc. are also presented.

References