Concepedia

Publication | Closed Access

Evaluation of Optic Disc Measurements with the Glaucoma-Scope

11

Citations

0

References

1996

Year

Abstract

The inter- and intraobserver reliability and low variation of the Glaucoma-Scope readings indicates moderate to substantial reproducibility for detecting changes in optic disc's topography under selected conditions. However, eyes with moderate cataract, aphakia, pseudophakia, mild corneal opacity, darkly pigmented fundi or myopia were inconsistently analyzable. In a small group of randomly selected eyes there was very low agreement between the glaucoma expert's evaluation of the optic disc and that of the Glaucoma-Scope. These findings raise questions about the usefulness of the instrument as a diagnostic tool for glaucoma but show promise for detecting change when good images are possible.