Publication | Closed Access
Calibration of the Gamma Knife Perfexion using TG‐21 and the solid water Leksell dosimetry phantom
27
Citations
18
References
2011
Year
The TG-21 dosimetry protocol, performed in a solid water phantom in conjunction with modern dosimeters and phantom material and chamber-dependent factors, can yield an accurate dose measurement in the unique GK treatment geometry. The technique described here can be easily adopted by institutions worldwide since all equipment and instruments used are commercially available, thus reducing the existing interinstitutional variation in GK calibration techniques. Relative output factor measurements made in this same solid water phantom were used to verify the relative output factors provided by Elekta and agreed excellently with output factors expected to be released in conjunction with GAMMAPLAN version 10.
| Year | Citations | |
|---|---|---|
Page 1
Page 1