Abstract

The mean defibrillation energy requirement does not change significantly after 2 years of biphasic nonthoracotomy defibrillator system implantation. However, approximately 15% of patients develop a 10-J or greater elevation in the defibrillation energy requirement, and 3% may require a defibrillation system revision. Therefore, a yearly evaluation of the defibrillation energy requirement may be appropriate.