The Analyst · 1991 · 19 citations · 2 references
PharmacyChemistryAnalytical ChemistryElemental AnalysisMass Spectrometric AnalysisElemental Composition AnalysisHeavy Metal AnalysisInorganic Elemental AnalysisMass Spectrometric Elemental AnalysisMass SpectrometryTrace Element AnalysisAnalytical Detection LimitsInternal Standard AnalysisMass Spectrometric QuantificationMass Spectrometry Of NuclidesUltra-trace AnalysisUltra-trace Element AnalysisUltra-trace Element DeterminationUranium Measurement
A sensitive method is decribed for the simultaneous determination of ultra-trace amounts of uranium and thorium in high-purity aluminium by inductively coupled plasma mass spectrometry (ICP-MS). Uranium and thorium were separated from a sample solution of 10 mol dm–3 hydrochloric acid by extraction with a 10% v/v solution of tributyl phosphate in cyclohexane. The internal standard method was used for quantification by ICP-MS. For a sample mass of 10 g, the detection limits for uranium and thorium are 7 and 8 pg g–1, respectively.
2