Analytical Chemistry · 1993 · 76 citations · 10 references
EngineeringBiochemistryBioengineeringBiomaterialsBiosensorsElectroanalytical MethodsBioanalytical ChemistryBiological MembranesBiosensor DesignElectrochemical BiosensorsEnzyme-based BiosensorsMembrane Self-assemblyBiotin-streptavidin ImmobilizationGlucose BiosensingGlucose BiosensorsGlucose OxidaseGlucose SensingGlucose SensorsOxidase-based Biosensors
A simple and fast procedure to prepare a glucose-sensitive minielectrode is presented. It is based on a biotin-modified phospholipid bilayer to which a streptavidin-glucose oxidase complex is coupled. The assay is based on the electrochemical detection of enzymatically generated hydrogen peroxide at the potential +670 mV. In the case of an air-saturated buffering solution the response to glucose was measured up to 50 mmol.L-1, with a linear portion up to 7 mmol.L-1. The influence of oxygen tension, pH, and temperature as well as possibly interfering substances was investigated. The prospect usage for the measurement of blood and urine was tested.
10